I've been having a lot of trouble with the variation in strip readings at each test time. I'm literally having to go through 3-4 strips each test to get a number I think might be right.
For example (and this is happening multiple times a day)...at +3 today I got 223, 76, 74, 201. His morning numbers were similar (484 and 495) so any of these numbers +3 would have been a huge drop and I settled on the 201? At least this one had two and two in the same family of numbers. One test yesterday had 430, 130, 306...
So far the diagnostics I've done:
-Changed the meter battery
-Did a control solution test (which looked completely fine)
-Verified that the code on the strips matched the code on the meter
At this rate, I'm blowing through strips and don't know if I'm recording accurate numbers. Has anyone else had this kind of trouble with the Alpha Trak meter?
For example (and this is happening multiple times a day)...at +3 today I got 223, 76, 74, 201. His morning numbers were similar (484 and 495) so any of these numbers +3 would have been a huge drop and I settled on the 201? At least this one had two and two in the same family of numbers. One test yesterday had 430, 130, 306...
So far the diagnostics I've done:
-Changed the meter battery
-Did a control solution test (which looked completely fine)
-Verified that the code on the strips matched the code on the meter
At this rate, I'm blowing through strips and don't know if I'm recording accurate numbers. Has anyone else had this kind of trouble with the Alpha Trak meter?
I'll tag a couple members that use AT2 more often